MPM700
The MPM700 is an Android based Patient Monitor with a 7" touch screen display. It has a built-in web server that makes it an IoT capable patient monitor that can be accessed by a web browser from anywhere.
- EKG (5 leads)
 - Blood Pressure
 - Pulse/SPO2
 - Temperature
 - Weight (Bluetooth connected)
 - Blood Glucose Reader
 - Built in 110-200v power supply
 - USB input jack for any USB device
 - USB output data jack.
 - Electronic Patient Record
 

Price
USD 699.00 (includes 1 set of adult size accessories)
Specifications
ECG
- 
measurement range: ±0.5mVp ~ ±5mVp
 - 
Heart Rate display range: 20bpm ~ 300bpm
 - 
Heart Rate display accuracy: ±2% or ±2bpm, whichever is greater.
 - 
Alarm delay time: <=12s
 - 
Input noise level for ECG amplifier: <=30uVp-p
 - 
Input impedance: >= 5M ohm (single ended)
 - 
CMRR (common mode rejection ratio): >=89dB
 - 
ECG bandwidth: 0.5~40Hz(monitoring mode), 0.05~75Hz(diagnostic mode)
 
SPO2
- SpO2 measuring range: 35% ~ 99%
 - SpO2 measuring accuracy: Arms <=3% for SpO2 range from 70% to 100%
 - Pulse Rate measurement: 30~240bpm with 2% or ±2bpm accuracy
 - Update rate: 8 pulse beats
 - Low perfusion performance: Accuracy of SpO2 & PR still meets the declared
 - specification while the pulse amplitude modulation ratio is as low as 0.4%.
 
NIBP
- Pneumatic pressure measurement: 0~300mmHg with ±3mmHg accuracy
 - Air leakage: <2mmHg/minute @ 300mmHg with 500cc cuff volume
 - Systolic pressure range: 40~270mmHg for adult, 40~200mmHg for pediatric
 - Diastolic pressure range: 10~210mmHg for adult, 10~150mmHg for pediatric
 - MAP range: 20~230mmHg for adult, 20~165mmHg for pediatric
 - BP accuracy: Mean error <= ±5mmHg, Standard deviation <= 8mmHg.
 - Pulse Rate measurement: 40~240bpm with 2% or ±2bpm accuracy
 - Blood pressure measuring time:< 30 sec. for typical adult (80/120mmHg, 75bpm) at manual or auto mode;
 - Time interval selection at auto mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90 minutes.
 - Over pressure protection with limit of 320±10mmHg by dual transducers
 
Respiration
- Respiratory Rate measurement range: 0~120rpm
 - Respiratory Rate measurement accuracy: ±10%
 
Temperature
- Temperature measurement range: 25.0~45.0 °C
 - Temperature measurement accuracy: ±0.2 °C
 - response time: <=30 sec.
 
Blood Sugar
- measurement range from 20-600 ml/dl
 - 0.7μL Blood sample requirement
 - used strip detection
 - response time < 10secs
 
Body Weight
- measurement range from 5kg-180kg
 - BMI Caliculations
 - Measurement in Lbs and Kgs.
 - Bluetooth enabled
 
Communication Support:
- 3G/GPRS packet data
 - Wifi (802.11 a/b/g)
 - USB connectivity to PC/Laptop
 
Display and Control Device:
- Android Tablet
 - Laptop/PC
 - Mobile Phone
 
 
              
              Next page: MPM 1300